Main Specifications
The module is designed for measuring, processing and transmitting of the following parameters:
- temperature (2 measurement channels of infrared sensors + 1 channel of thermal resistance sensor;
- invasive arterial blood pressure (4 measurement channels);
- flowmeter and bubble detector (2 measurement channels);
- pH factor (1 measurement channel);
- saturation (SpO2) (1 measurement channel).
Functionality
The OEM module provides:
- continuous measurement of the preset parameters through the measurement channels;
- data processing of the measurement channels;
- sensor connection control;
- sensor state control;
- interaction (request reception, data transmission) by means of the communication channel.
Measurement channels
Flowmeter (of water, physiological solution, custodiol, blood):
- Range, l/min: 0,1 – 1,6. Error (not less than), ml/min: ± 32.
- Range, l/min: 1,6 - 10. Error (not less than): ± 2 % of measured value.
Temperature channel
- Range, °C: 0 – 60
- Error (not less than), °C: ± 0,5°C
Invasive arterial blood pressure channel
- Range, mm Hg.: -99 – 350
- Error (not less than), %: ± 5
Blood oxygen saturation channel:
- Range, %: 70 – 100. Error (not less than), %: ± 3.
- Range, %: 10 – 70. Error is not defined.
pH factor in blood measurement channel:
- Range: 1-12
- Error (not less than): ±0,05
Power supply
- Voltage, V: 12 ± 5%
- Load current (not less than), A: 0,3
IR temperature sensor
Infrared temperature sensor OTFADR3801.
Designed for medical tubes with diameter of 3/8”.
SpO2 sensor
Magistral SpO2 sensor OSTADR3801.
Designed for medical tubes with diameter of 3/8”.
Flowmeter sensor and bubble detector
Flowmeter sensor and bubble detector.
Designed for medical tubes with diameter of 3/8”.
ЗАПОЛНИТЕ ВИДЫ ИСПОЛНЕНИЯ