Purpose
- Emergency management of blood circulation and oxygenation of blood in conditions of resuscitation, medicine of catastrophes, cardiosurgery and transplantation
- Emergency management of blood circulation in the organs of the donor with the purpose of their resuscitation at the pre-implantation stage ("block", inside the donor's body) for selection and maintenance of viability at all stages of organ transplantation from the donor to the recipient
- Treatment of patients with acute cardiorespiratory insufficiency of various etiologies, both in hospital and during transportation
- For the period of waiting for the donor organ and stabilization of the patient's condition for urgent patients who are on the waiting list for a heart transplant or a heart-lung system
- Isolated chemo-perfusion for the treatment of cancer patients
Advantages
- Automatic and manual perfusion control
- Portability: weight – 2.47 kg, size –270 х 120 х 230 mm
- Autonomous power supply for at least 5 hours
- Remote on-line monitoring and perfusion results database
- Efficiency of device preparation for operation (15-20 min), Simplicity and convenience of adjustment and control
- Modular design, involving the connection of new measurement channels
Characteristics
- AC power supply 220±10 V, 50 Hz
- Number of measuring channels: pressure - 2, volumetric flow - 1, temperature -1; blood saturation - 1
- Adjustable volumetric flow rate of perfusate: 0,3-10 l/min
- Rotational speed range of the pump rotor: 300 - 4500 rpm
- The maximum pressure at the pump outlet – not less than 250 mm Hg
- Pump type - centrifugal
- Degree of protection from penetration of water and dust at least IP54 for field operation and for transportation by road/air
- Perfusion Index Control
- Pulsing/laminar modes of perfusion
- Keypad lock for prevention of unintentional actions
- Built-in automatic calibration of measurement channels
- Built-in alarm/warning system
The system was developed by
The device was developed by OOO "BIOSOFT-M" (design, hardware and software) in partnership with OOO "TransBioTech" (method and medical aspects of clinical implementation) and sponsored by the "Skolkovo" Fund in year 2017.

OOO "TransBioTech"

IR temperature sensor
Infrared temperature sensor OTFADR3801.
Designed for medical tubes with diameter of 3/8”.
SpO2 sensor
Magistral SpO2 sensor OSTADR3801.
Designed for medical tubes with diameter of 3/8”.